16 ProLink
®
II Software for Micro Motion
®
Transmitters
Installation and Setup
Figure 2-8 Modbus/RS-485 temporary connection to the service port
Figure 2-9 HART/Bell 202 temporary connection to HART clips
Attach serial port or USB adapter
if necessary (not shown)
Service port terminals
Transmitter
RS-232 <-> 2-wire RS-485
Interface Converter
Notes:
For Model 1700/2700, Model 2400S, and LF-Series field-mount transmitters, the service port is available at any
time.
For Model 1500/2500, Series 3000, and LF-Series DIN rail transmitters, the service port is available during a
10-second interval after power-up. If a service port connection is made during this time, the ports will remain in
service port mode.
Ensure that no other Modbus master devices are currently active on the network.
All service ports are accessed using the default address of 111. If you are connecting over a multidrop network
with multiple service ports, it is not possible to specify which device to connect to.
VIATOR
HART clips
Bell 202 signal
converter
Notes:
If necessary, add resistance across the HART clips to bring the total resistance within 250–600 Ω.
Zobrazit stránku 21