Micro Motion
®
2-Wire Coriolis Flow and Density Meters 3
Liquid flow performance
Mass Volume
(1)
(1) Specifications for volumetric flow rate are based on a process-fluid density of 1000 kg/m
3
. For fluids with density other than
1000 kg/m
3
, the volumetric flow rate equals the mass flow rate divided by the fluid’s density.
kg/h l/h
Maximum flow rate CMF025 2180 2180
CMF050 6800 6800
CMF100 27200 27200
CMF200 87100 87100
Mass and volume flow accuracy
(2)
(2) Stated flow accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on
reference conditions of water at 20 to 25 °C and 1 to 2 bar, unless otherwise noted.
±0,10% of rate
(3)
(3) When flow rate is less than zero stability / 0,001, accuracy = ±[(zero stability / flow rate) × 100]% of rate and
repeatability = ±[½(zero stability / flow rate) × 100]% of rate.
Mass and volume flow repeatability ±0,05% of rate
(3)
kg/h
Zero stability CMF025 0,027
CMF050 0,163
CMF100 0,680
CMF200 2,18
Typical accuracy and turndown
The graph below is an example of the relationship between accuracy and turndown on water. To determine actual accuracy and
turndown with your process variables, use the Micro Motion product selector, available at www.micromotion.com.
–2.5
–2.0
–1.5
–1.0
–0.5
0
0.5
1.0
1.5
2.0
2.5
1009080706050403 020100
100:1
20:1
10:1
1:1
Flow rate, % of maximum
Accuracy, %
Turndown 500:1 100:1 20:1 10:1 1:1
Accuracy (±%) 2,40 0,50 0,10 0,10 0,10
2.5
2.0
1.5
1.0
0.5
0
–2.5
–2.0
–1.5
–1.0
–0.5
0 10 20 30 40 50 60 70 80 90 100
Komentáře k této Příručce