Model Solu Comp®IIDual Input pH/Conductivity AnalyzerInstruction ManualPN 51-1055pHC/rev.KFebruary 2006
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0. DESCRIPTION AND SPECIFICATIONS1.1 FEATURES AND APPLICATIONS1.2 SPECIFICAT
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS21.2 SPECIFICATIONS - GeneralCase: ABS (panel-mount), polycarbonate (pipe- andsurface-mou
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSMeasures conductivity in the range 0 to 20,000 µS/cm.Display choices are conductivity, res
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS4When used with Model Series 200 ToroidalConductivity Sensors, display choices are conduct
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSFor use with any standard pH or ORP sensor and allUniloc sensors and junction boxes with b
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONSWhen used with a chlorine specific membrane-coveredamperometric sensor, display choices ar
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS7When used with an oxygen permeable membrane-covered amperometric sensor, display choices
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS81.3 ORDERING INFORMATIONThe Solu Comp II analyzers offer the choice of single or dual sen
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS9COMBINATION GUIDE55* For D.O. sensors with 22k thermistor, use Suite 1 only.
MODEL SOLU COMP II SECTION 1.0DESCRIPTION AND SPECIFICATIONS10ACCESSORIES (Weights are rounded up to nearest whole lb or 0.5 kg)PART NUMBER DESCRIPTIO
ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!Your purchase from Rosemount Analytical, Inc. hasresulted in one of the finest instruments ava
11SECTION 2.0. INSTALLATIONMODEL SOLU COMP II SECTION 2.0INSTALLATION2.1 UNPACKING AND INSPECTION2.2 INSTALLATIONType of Mounting SectionPanel 2.2.2
12MODEL SOLU COMP II SECTION 2.0INSTALLATIONFIGURE 2-1. Panel Mount InstallationAccess to the wiring terminals is through the rear cover. Four screws
MODEL SOLU COMP II SECTION 2.0INSTALLATIONFIGURE 2-2. Pipe Mount InstallationThe front panel is hinged at the bottom. The panel swings down for access
MODEL SOLU COMP II SECTION 2.0INSTALLATION14FIGURE 2-4. Surface Mount InstallationThe front panel is hinged at the bottom. The panel swings down for a
3.2 POWER, ALARM, OUTPUT, AND SENSOR CONNECTIONSThe Solu Comp II is available in two mounting configurations.The positions of the power, alarm, outpu
16MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-2. Wiring Connections for Solu Comp II Model 1055-01-10 (Panel Mount with 1
MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-5. Wiring Connections for Solu Comp II Model 1055-02-11 (Surface/Pipe Mountin
18MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-6. Panel Mount Wiring to Models 399-09-62, 400, 401, 402, 403, and 404.DWG. NO. REV.41055105 B
MODEL SOLU COMP II SECTION 3.0WIRING19FIGURE 3-7. Pipe/Wall Mount Wiring to Models 399-14, 400, 401, 402, 403, and 404.DWG. NO. REV.41055106 B
20MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-8. Pipe/Wall Mount Wiring to Models 399VP-09, 400, 401, 402, 403, and 404.DWG. NO. REV.41055107 B
QUICK START GUIDEFOR MODEL SOLU COMP II pH/CONDUCTIVITY ANALYZER(Model Option 1055-22-30)1. Refer to Section 2.0 for installation instructions.2. Wi
MODEL SOLU COMP II SECTION 3.0WIRING21FIGURE 3-9. Panel Mount Wiring to Models 399VP-09, 400, 401, 402, 403, and 404.DWG. NO. REV.41055108 B
22MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-10. Pipe/Wall Mount Wiring to Models 399-09-62, 400, 401, 402, 403, and 404.DWG. NO. REV.41055109 B
MODEL SOLU COMP II SECTION 3.0WIRING23FIGURE 3-11. Panel Mount Wiring to Models 399-14, 400, 401, 402, 403, and 404.DWG. NO. REV.41055110 B
24MODEL SOLU COMP II SECTION 3.0WIRINGFIGURE 3-12. Wiring Model 1055-11-22-32 to Models 399-09-62 and GP-1.DWG. NO. REV.40105536 C
MODEL SOLU COMP II SECTION 3.0WIRING25FIGURE 3-13. Wiring Model 1055-10-22-32 to Model 399-14DWG. NO. REV.40105559 EFIGURE 3-14. Wiring Model 1055 t
26MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATIONSECTION 4.0DISPLAY AND OPERATION4.1 DISPLAY4.2 KEYPAD4.3 PROGRAMMING AND CALIBRATING THE SOLU
27MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATION4.3 PROGRAMMING AND CALIBRATING THE SOLU COMP II- TUTORIALSetting up and calibrating the Solu Co
MODEL SOLU COMP II SECTION 4.0DISPLAY AND OPERATION1. If a security code has been programmed, pressing MENU causes thesecurity screen to appear.2. Ent
29MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERSECTION 5.0PROGRAMMING THE ANALYZER5.1 GENERAL5.2 CHANGING STARTUP SETTINGS5.3 CONFIGURING
30MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGS1. SENSOR-OUTPUT ASSIGNMENTS (pH, ORP, redox, conductivity, resist
5. Choose the desired language. Select >> to show more choices.6. Choose the number of sensors wired to the analyzer. NOTEIf One sensor is chose
MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGS (continued)6. MISCELLANEOUS SETTINGSSectionLanguage English 5.11Hol
32MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.3 CONFIGURING AND RANGING THE OUTPUTS.5.3.1 Purpose The Solu Comp II accepts input from a pH
MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.3.3. Procedure: Configure Outputs.To choose a menu item, move the cursor to the item and press
34Alarm relays are single pole-double throw (SPDT). When an alarm is activated, the coil is energized. When an alarm activates, AL1, AL2, or AL3 (as a
35MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.4.3 Procedure: Configuring AlarmsTo choose a menu item, move the cursor to the item and pres
36MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.4.4 Procedure: Programming Alarm SetpointsTo choose a menu item, move the cursor to the item
37MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5 SELECTING THE TYPE OF MEASUREMENT AND ACTIVATING SOLUTION TEMPERATURE CORRECTIONS5.5.1 Pu
MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5.3 Definitions — conductivity/resistivity1. NEUTRAL SALT CORRECTION. Neutral salt temperature
MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.5.4 Procedure.To choose a menu item, move the cursor to the item and press ENTER. To store a n
40MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.6 CHOOSING TEMPERATURE UNITS AND MANUAL OR AUTOMATIC TEMPERATURECOMPENSATION5.6.1 PurposeTh
QUICK REFERENCE GUIDEMENU TREE FOR pH/ORP/REDOX/CONDUCTIVITY/RESISTIVITY/TDS MEASUREMENTSpH/cond/temp
41MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.7 SETTING A SECURITY CODE5.7.1 Purpose.This section describes how to set a security code. Th
MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.8 NOISE REJECTION5.8.1 Purpose.For maximum noise rejection, the frequency of the ac power must
43MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.11 SELECTING A DEFAULT SCREEN, LANGUAGE, AND SCREEN CONTRAST5.11.1 PurposeThis section descr
44MODEL SOLU COMP II SECTION 5.0PROGRAMMING THE ANALYZER5.11.4 Procedure: Changing Screen Contrast1. Press MENU. The main menu screen appears. Choose
45MODEL SOLU COMP II SECTION 6.0CALIBRATIONSECTION 6.0CALIBRATION6.1 INTRODUCTION6.2 CALIBRATING TEMPERATURE6.3 AUTO CALIBRATION - pH6.4 MANUAL
46MODEL SOLU COMP II SECTION 6.0CALIBRATION6.2 CALIBRATING TEMPERATURE6.2.1 PurposeTemperature is important in the measurement of pH, ORP, and conduct
476.2.2 Procedure1. Remove the sensor from the process. Place it in an insulated container of water along with a calibrated ther-mometer. Submerge at
MODEL SOLU COMP II SECTION 6.0CALIBRATION6.3 AUTO CALIBRATION — pH6.3.1 Purpose1. New sensors must be calibrated before use. Regular recalibration is
49MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Calibrate.b. Choose Sensor1 (pH sensor). c. Choose Meas
50MODEL SOLU COMP II SECTION 6.0CALIBRATION6.4 MANUAL CALIBRATION — pH6.4.1 Purpose1. New sensors must be calibrated before use. Regular recalibration
iMODEL SOLU COMP II TABLE OF CONTENTSMODEL SOLU COMP IIDUAL INPUT pH/CONDUCTIVITY ANALYZERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECI
51MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Calibrate.b. Choose Sensor1 (pH sensor). c. Choose Meas
52MODEL SOLU COMP II SECTION 6.0CALIBRATION6.5 STANDARDIZATION — pH6.5.1 Purpose1. The pH measured by the Solu Comp II analyzer can be changed to mat
MODEL SOLU COMP II SECTION 6.0CALIBRATION6.6 ENTERING A KNOWN SLOPE VALUE — pH6.6.1 PurposeIf the electrode slope is known from other measurements, it
54MODEL SOLU COMP II SECTION 6.0CALIBRATION6.7 ORP CALIBRATION6.7.1 Purpose1. For process control, it is often important to make the measured ORP agre
6.8 CALIBRATING AN IN-SERVICE SENSOR — CONDUCTIVITY6.8.1 Purpose1. After a conductivity sensor has been in service for aperiod of time, recalibration
56MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Program.b. Choose Measurement.c. Choose Sensor2 (conduc
MODEL SOLU COMP II SECTION 6.0CALIBRATIONPossible Cal ErrorProceed?YYeessNoUpdated S2 CellConst: 1.0013/cmh. If the change is more than 10% of the pre
58MODEL SOLU COMP II SECTION 6.0CALIBRATIONa. Press MENU. The main menu screen appears. Choose Program.b. Choose Measurement.c. Choose Sensor2 (conduc
MODEL SOLU COMP II SECTION 6.0CALIBRATIONCalibrate?SSeennssoorr11Sensor2Cal Sensor2?MMeeaassuurreemmeennttTempCalS2? InProcessMeterCCeellllCCoonnssttC
60MODEL SOLU COMP II SECTION 6.0CALIBRATION6.10 CALIBRATING THE ANALYZER - CONDUCTIVITY.6.10.1 Purpose1. After a conductivity analyzer has been in ser
iiSection Title Page6.0 CALIBRATION ... 456.1 Int
MODEL SOLU COMP II SECTION 6.0CALIBRATION9. If a resistance value outside the range 100 Ω to 180 kΩ was entered, thecalibration will not be accepted.
62MODEL SOLU COMP II SECTION 7.0MAINTENANCESECTION 7.0MAINTENANCE7.1 OVERVIEW7.2 REPLACEMENT PARTS7.1 OVERVIEWThe Solu Comp II analyzer needs little
MODEL SOLU COMP II SECTION 7.0MAINTENANCE63Location in ShippingFigure 7-1 PN Description Weight1 23823-00 Panel mounting kit, includes four brackets a
64MODEL SOLU COMP II SECTION 7.0MAINTENANCELocation in ShippingFigure 7-2 PN Description Weight1 note Screw, 6-32 x 1.38 in.2 note Washer3 23834-00 En
65MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTINGSECTION 8.0TROUBLESHOOTING8.1 OVERVIEW8.2 TROUBLESHOOTING USING FAULT CODES8.3 TROUBLESHOOTING WHEN N
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.2.2 pH Sensitive Glass Membrane is BrokenThe Solu Comp II continuously measures the impedance between
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH.Problem See SectionNew temperature during cali
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3.3 Calibration Error during Standardization. During standardization, the millivolt signal from the pH
69MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.3.7 Calibration Was Successful, but Process pH Is Grossly Wrong and/or Noisy.Grossly wrong or noisy r
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.4 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — CONDUCTIVITY8.4.1 Calibration Error Using Standa
MODEL SOLU COMP II TABLE OF CONTENTSLIST OF FIGURESNumber Title Page2-1 Panel Mount Installation ...
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.4.5 Measurement seems wrong.A. If sensor is being installed for the first time, check the wiring conne
72MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.6 SIMULATING INPUTS — pH8.6.1 GeneralThis section describes how to simulate a pH input into the Solu
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTING8.7 SIMULATING INPUTS — CONDUCTIVITYTo check the performance of the analyzer, use standard resistors to
Temp. (°C) Pt 100 (ΩΩ) Pt 1000 (ΩΩ)0 100.0 100010 103.9 103920 107.8 107825 109.7 109730 111.7 111740 115.5 115550 119.4 119460 123.2 123270 127.1 127
MODEL SOLU COMP II SECTION 8.0TROUBLESHOOTINGFIGURE 8-7. Checking for a Poisoned ReferenceElectrode.Refer to the sensor wiring diagram to identify the
76MODEL SOLU COMP II SECTION 9.0RETURN OF MATERIALSECTION 9.0RETURN OF MATERIAL9.1 GENERAL.To expedite the repair and return of instruments, proper co
WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services pro
Credit Cards for U.S. Purchases Only.The right people,the right answers,right now.ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITEhttp://www.raihome.com
About This DocumentThis manual contains instructions for installation and operation of the Solu Comp IIModel 1055 Dual Input pH/Conductivity Analyzer.