Emerson Rosemount 2230 Uživatelský manuál Strana 72

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 80
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 71
Reference Manual
300560EN, Rev AA
May 2011
Rosemount 2230
B-4
Appendix B. Product Certifications
B.3.2 Factory Mutual
Canadian Approvals
Certificate of Compliance: 3037177C.
Figure B-2. Factory Mutual
Intrinsic Safety Canadian
Approval Label
I6
FISCO Field Device (Fieldbus Terminals)
Intrinsically Safe for Class I, II, III Division 1, Groups A, B, C, D, E, F and G
Temperature Class T4, Ambient Temperature Limits: -50 °C to +70 °C
Ui=17.5V, Ii=380mA, Pi=5.32W, Ci=2.1nF, Li=1.1µH
Entity (Fieldbus Terminals)
Intrinsically Safe for Class I, II, III Division 1, Groups A, B, C, D, E, F and G
Temperature Class T4, Ambient Temperature Limits: -50 °C to +70 °C
Ui=30V, Ii=300mA, Pi=1.3W, Ci=2.1nF, Li=1.1µH
Install per Control Drawing 9240040-949
Special Conditions of Use
1. The associated apparatus through which the equipment is supplied shall
provide galvanic isolation between the input and output of the associated
apparatus.
Zobrazit stránku 71
1 2 ... 67 68 69 70 71 72 73 74 75 76 77 78 79 80

Komentáře k této Příručce

Žádné komentáře