Product Data Sheet
00813-0100-4530, Rev AA
Catalog 2008 - 2009
Rosemount 5300 Series
Level-8
Flange Rating
•ANSI:
According to ANSI B16.5 Table 2-2.3.
Standard: Max. 302 °F/580 psig
(150 °C/40 Bar).
HP/HTHP: Up to Class 2500.
•EN:
According to EN 1092-1 Table 18, material
group 13E0.
Standard: Max. 302 °F/580 psig
(150 °C/40 Bar).
HP/HTHP: Up to PN 320.
• Fisher & Masoneilan:
According to ANSI B16.5 Table 2-2.3.
Standard: Max. 302 °F/580 psig
(150 °C/40 Bar).
HP/HTHP: Up to Class 600.
• JIS:
According to JIS B2220 Table 2.3
Standard: 10K/20K/150C.
HP: 10K/20K/200C.
HTHP: 10K/20K/400C
Ambient Temperature
The maximum ambient temperature depends on the
process temperature according to the graph below.
Nozzle insulation for the HTHP version should not
exceed 4 in. (10 cm).
PROBES
Several versions of the probes are available: Coaxial
(perforated and non-perforated versions), Rigid Twin
and Rigid Single Lead, Flexible Twin and Flexible
Single Lead. Probes can be ordered in different
materials, and there are options for extreme
temperatures and pressure.
Total probe length is defined from the upper
reference point to the end of the probe (weight
included if applicable).
For guidance in probe selection, see page 10.
The table on page 9 shows what probe types that are
available for different materials of construction and
for the HP & HTHP options.
Process
Temperature
°F (°C)
Ambient Temperature
°F (°C)
185 (85)
131 (55)
100 (38)
50 (10)
0 (-18)
0 (-18) 200 (93) 400 (204) 600 (316)
392 (200)
752 (400)
Total Probe Length and Upper Reference Point (right
below flange / thread).
Upper
Reference
Point
Total Probe
Length
NPT
BSP/G
Flange