Series 3000 Transmitters and Discrete Controllers 9
Model 3500/3700 performance specifications
Sensor model Flow accuracy*
ELITE
®
liquid ± 0,10% ± [(zero stability/flow rate) x 100]% of rate
gas ± 0,50% ± [(zero stability/flow rate) x 100]% of rate
T-Series liquid only ± 0,15% ± [(zero stability/flow rate) x 100]% of rate
F-Series liquid ± 0,20% ± [(zero stability/flow rate) x 100]% of rate
gas ± 0,70% ± [(zero stability/flow rate) x 100]% of rate
D (except DH38), DL and DT liquid ± 0,15% ± [(zero stability/flow rate) x 100]% of rate
gas ± 0,65% ± [(zero stability/flow rate) x 100]% of rate
DH38 liquid ± 0,15% ± [(zero stability/flow rate) x 100]% of rate
gas ± 0,50% ± [(zero stability/flow rate) x 100]% of rate
Sensor model Flow repeatability*
ELITE liquid ± 0,05% ± [½(zero stability/flow rate) x 100]% of rate
gas ± 0,25% ± [(zero stability/flow rate) x 100]% of rate
T-Series liquid only ± 0,05% ± [½(zero stability/flow rate) x 100]% of rate
F-Series liquid ± 0,10% ± [½(zero stability/flow rate) x 100]% of rate
gas ± 0,35% ± [(zero stability/flow rate) x 100]% of rate
D (except DH38), DL and DT liquid ± 0,05% ± [½(zero stability/flow rate) x 100]% of rate
gas ± 0,30% ± [(zero stability/flow rate) x 100]% of rate
DH38 liquid ± 0,05% ± [½(zero stability/flow rate) x 100]% of rate
gas ± 0,25% ± [½(zero stability/flow rate) x 100]% of rate
Sensor model Density accuracy Density repeatability
ELITE (except high-pressure CMF010P) liquid ± 0,5 kg/m
3
± 0,2 kg/m
3
gas ± 2,0 kg/m
3
± 1,0 kg/m
3
ELITE high-pressure CMF010P liquid ± 2,0 kg/m
3
± 1,0 kg/m
3
gas ± 8,0 kg/m
3
± 4,0 kg/m
3
T-Series liquid only ± 2,0 kg/m
3
± 0,5 kg/m
3
F-Series liquid only ± 2,0 kg/m
3
± 1,0 kg/m
3
D6, D12, D25, D40, liquid only ± 2,0 kg/m
3
± 1,0 kg/m
3
DH100, DH150
DH6, DH12, DH25, DH38 liquid only ± 4,0 kg/m
3
± 2,0 kg/m
3
D65, DL65, DT65, liquid only ± 1,0 kg/m
3
± 0,5 kg/m
3
D100, DT100, D150, DT150,
DH300
DL100, DL200 liquid only ± 0,5 kg/m
3
± 0,2 kg/m
3
D300, D600
Sensor model Temperature accuracy Temperature repeatability
All models ± 1°C ± 0,5% of reading in °C ± 0,2°C
* Flow accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications are based on reference conditions of water at
20 to 25°C, and 1 to 2 bar unless otherwise noted. For values of zero stability, refer to product specifications for each sensor.