Reference Manual
00809-0100-4801, Rev FA
October 2010
5-7
Rosemount 3051S Series
Table 5-2. Bolt Installation
Torque Values
4. If you replaced the PTFE SuperModule O-rings, re-torque the flange
bolts after installation to compensate for cold flow.
5. Install the drain/vent valve.
a. Apply sealing tape to the threads on the seat. Starting at the base of
the valve with the threaded end pointing toward the installer, apply
two clockwise turns of sealing tape.
b. Take care to place the opening on the valve so that process fluid will
drain toward the ground and away from human contact when the
valve is opened.
c. Tighten the drain/vent valve to 250 in-lb. (28.25 N-m).
NOTE
After replacing O-rings on Range 1 transmitters and re-installing the process
flange, expose the transmitter to a temperature of 185 °F (85 °C) for two
hours. Then re-tighten the flange bolts in a cross pattern, and again expose
the transmitter to a temperature of 185 °F (85 °C) for two hours before
calibration.
Bolt Material Initial Torque Value Final Torque Value
CS-ASTM-A445 Standard 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
316 SST—Option L4 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)
ASTM-A-193-B7M—Option L5 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
Alloy K-500 —Option L6 300 in-lb. (34 N-m) 650 in-lb. (73 N-m)
ASTM-A-453-660—Option L7 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)
ASTM-A-193-B8M—Option L8 150 in-lb. (17 N-m) 300 in-lb. (34 N-m)