B-2
EN
Rosemount Inc
Factory Mutual (FM) Approvals
E5 Explosion-Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition
Proof for Class II, Division 1, Groups E, F, and G. Dust-Ignition Proof for
Class III, Division 1 hazardous locations when installed in accordance with
Rosemount Drawing 00644-1049. Non-Incendive for Class I, Division 2,
Groups A, B, C, and D. Ambient temperature limits: –50 to 85 °C. Conduit
seal not required for compliance with NEC 501-5a(1). Temperature Code T5
(T
amb
= 85 °C)
NOTE:
Approval E5 is only available with 644H option codes J5 and J6.
I5 Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F, G;
Non-incendive for Class I, Division 2, Groups A, B, C, D. Suitable for indoor
and outdoor locations NEMA 4x. Ambient temperature limits: –50 to 80 °C
when installed in accordance with Rosemount Drawing 00644-0009. FM
Entity Parameters: See installation drawing 00644-0009. Temperature Code
T5 (T
amb
= 80 °C) (T6 (T
amb
= 40 °C))
Special Conditions for Safe Use (X):
When the ouput of the associated apparatus does not exceed P
o
= 0.6 Watts,
the temperature code is T6 (T
amb
= 50 °C)
K5 Combination of E5 and I5
NOTE:
Approval K5 is only available with Model 644H option codes J5 and J6.
Canadian Standards Association (CSA) Approvals
I6 Intrinsically safe for Class I, Division 1, Groups A, B, C, and D when
connected in accordance with Rosemount drawing 00644-1064
C6 Combination of I6 and the following: Explosion-Proof for Class I, Division 1,
Groups B, C, and D. Dust-ignition proof for Class II, Division 1, Groups E, F,
and G. Dust-ignition proof for Class III, Division 1 hazardous locations when
installed in accordance with Rosemount Drawing 00644-1059. Suitable for
Class I, Division 2, Groups A, B, C, and D.
NOTE:
Approval C6 is only available with Model 644H option code J6.