Emerson ROSEMOUNT 5300 Specifikace Strana 52

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 106
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 51
Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
4-8
Figure 4-8. Echo Curve plot
indicating that the amplitude
threshold for the interface peak
is too high
If the Interface Threshold is too high, the signal amplitude peak at the
interface between the upper and lower products is not detected.
Recommended Actions
Lower Interface Threshold.
Strong fixed echo - Bent probe / Probe contacting Nozzle
A bent probe in contact with metal creates a strong echo and
measurement will lock onto the strong echo. Note that contamination also
may create strong disturbance echoes.
Figure 4-9. Measurement data
using a bent probe from refining
best practices
3.0 5.0
The Interface Threshold is above
the interface peak
Amplitude, mV
Distance, m
Interface Threshold
Surface Threshold
Reference Threshold
Zobrazit stránku 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 105 106

Komentáře k této Příručce

Žádné komentáře