Emerson 3051S Uživatelský manuál Strana 48

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 84
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 47
48
Rosemount 3051
September 2014
www.rosemount.com
Mounting position effects
Vibration effect
Less than ±0.1% of URL when tested per the requirements of IEC60770-1: 1999 field or pipeline with high vibration level
(10-60 Hz 0.21 mm displacement peak amplitude / 60-2000 Hz 3g).
Power supply effect
Less than ±0.005% of calibrated span per volt change.
Electromagnetic compatibility (EMC)
Transient protection (option code T1)
Tested in accordance with IEEE C62.41.2-2002, Location Category B
6 kV crest (0.5 s - 100 kHz)
3 kA crest (8 × 20 s)
6 kV crest (1.2 × 50 s)
Models Mounting position effects
3051C Zero shifts up to ±1.25 inH
2
O (3,11 mbar), which can be calibrated out. No span effect.
3051CA, 3051T Zero shifts up to ±2.5 inH
2
O (6,22 mbar), which can be calibrated out. No span effect.
3051L
With liquid level diaphragm in vertical plane, zero shift of up to ±1 inH
2
O (2,49 mbar). With
diaphragm in horizontal plane, zero shift of up to ±5 inH
2
O (12,43 mbar) plus extension length on
extended units. All zero shifts can be calibrated out. No span effect.
Meets all relevant requirements of EN61326-1:2006 and Namur NE-21.
(1)
(1) NAMUR NE-21 does not apply to wireless output code X.
Zobrazit stránku 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 83 84

Komentáře k této Příručce

Žádné komentáře