May 2015
20
Quick Start Guide
Note
Transmitters marked with NI CL 1, DIV 2 can be installed in Division 2 locations using general
Division 2 wiring methods or Nonincendive Field Wiring (NIFW). See Drawing 03151-1006.
IE FM FISCO
Certificate: 3012350
Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, FM Class 3611 – 2004,
FM Class 3810 – 2005, NEMA 250 – 2003
Markings: IS CL I, DIV 1, GP A, B, C, D; T4(-50 °C
≤ T
a
≤ +60 °C); when connected
per Rosemount drawing 03151-1006; Type 4X
Special Condition for Safe Use:
1. The Model 3051S/3051S-ERS Pressure Transmitter contains aluminum and is
considered to constitute a potential risk of ignition by impact or friction. Care
must be taken into account during installation and use to prevent impact and
friction.
Canada
E6 CSA Explosionproof, Dust-Ignitionproof, and Division 2
Certificate: 1143113
Standards: CAN/CSA C22.2 No. 0-10, CSA Std C22.2 No. 25-1966, CSA Std C22.2
No. 30-M1986, CAN/CSA C22.2 No. 94-M91, CSA Std C22.2 No.
142-M1987, CSA Std C22.2 No. 213-M1987, ANSI/ISA 12.27.01-2003,
CSA Std C22.2 No. 60529:05
Markings: Explosionproof Class I, Division 1, Groups B, C, D; Dust-Ignitionproof
Class II, Division 1, Groups E, F, G; Class III; suitable for Class I, Zone 1,
Group IIB+H2, T5; suitable for Class I, Division 2, Groups A, B, C, D;
suitable for Class I, Zone 2, Group IIC, T5; when connected per
Rosemount drawing 03151-1013; Type 4X
I6 CSA Intrinsically Safe
Certificate: 1143113
Standards: CAN/CSA C22.2 No. 0-10, CSA Std C22.2 No. 30-M1986, CAN/CSA
C22.2 No. 94-M91, CSA Std C22.2 No. 142-M1987, CSA Std C22.2 No.
157-92, ANSI/ISA 12.27.01-2003, CSA Std C22.2 No. 60529:05
Markings: Intrinsically Safe Class I, Division 1; suitable for Class 1, Zone 0, IIC,
T3C; when connected per Rosemount drawing 03151-1016; Type 4X
IF CSA FISCO
Certificate: 1143113
Standards: CAN/CSA C22.2 No. 0-10, CSA Std C22.2 No. 30-M1986, CAN/CSA
C22.2 No. 94-M91, CSA Std C22.2 No. 142-M1987, CSA Std C22.2
No. 157-92, ANSI/ISA 12.27.01-2003, CSA Std C22.2 No. 60529:05
Markings: FISCO Intrinsically Safe Class I, Division 1; suitable for Class I, Zone 0;
T3C; when installed per Rosemount drawing 03151-1016; Type 4X
4801_QIG_RevMA.fm Page 20 Wednesday, May 6, 2015 1:15 PM