Recall factory trim
The Recall Factory Trim button will restore the transmitter to the original factory characteriza-
tion curve. The Recall Factory Trim button can be useful for recovering from an inadvertent zero
trim or inaccurate pressure source.
When the recall factory trim function is used, the transmitter’s upper and lower trim values are
set to the values configured at the factory. If custom trim values were specified when the
transmitter was ordered, the device will recall those values. If custom trim values were not
specified, the device will recall the upper and lower sensor limits.
Last static pressure sensor trim
The current upper and lower trim points can be seen under the Last Static Pressure Sensor Trim
Points heading.
Calibration type
The calibration type drop-down menu allows the user to note the type of device last used to
calibrate the sensor (either Differential, Gage, or Absolute). This field does not affect the
calibration of the device.
4.3.5 Process temperature sensor calibration
The Temperature Calibration tab allows the user to perform a sensor trim and configure the
sensor matching of a process temperature sensor, see Figure 4-4.
Figure 4-4. Calibration - temperature calibration tab
Mass and Energy Flow Fast Keys 1, 2, 5, 5
Direct Process Var. Output Fast Keys 1, 2, 4, 5