Reference Manual
00809-0100-4803, Rev DA
Section 3: Configuration
April 2013
Configuration
Energy and time conversion example:
To find the Base per Custom relationship for a custom unit of thousands of BTU per day (kBtud)
and a base unit of BTU per hour (Btu/h), input the following:
Custom Unit = kBtud
Base Unit = Btu/h
Because:
1 kBtu (Thousands of BTU)= 1000 Btu and
1 d (Day) = 24 h (Hours)
Then:
1 kBtud = x x = 41.6667 Btu/h
1 kBtud = 41.6667 Btu/h
Therefore:
Base per Custom = = = 41.6667
Figure 3-31. Energy rate custom units - Energy and time conversion example
3. Under the Low Flow Cutoff heading, edit the current Minimum DP Value as needed. The
unit for this value is the user-selected DP unit. If the measured DP value is less than the
minimum DP value, the transmitter will calculate the energy value to be zero.
4. The Sensor Limits and Minimum Span can be viewed under the Energy Rate Sensor Limits
heading.
1kBtu
1d
---------------------
1000 Btu
1kBtu
---------------------------
1d
24 h
--------------
Number of Base Units
1xCustomxUnit
-----------------------------------------------------------------------
41.6667 Btu h
1kBtud
-------------------------------------------