162 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Calibrating the Transmitter continued
Figure C-3 D3 or D3 and D4 density calibration – ProLink II
Fill sensor with D3 fluid
Calibration
Density
Density cal - Point 3
Do Cal
Enter density of D3 fluid
Fill sensor with D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density
Density cal - Point 4
Do Cal
Enter density of D4 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Exit
Close Close