Emerson MICRO MOTION 2700 Instalační manuál Strana 55

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 264
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 54
Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 45
Required Transmitter Configuration continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Figure 6-1 Sample calibration tags – All sensors except T-Series
Figure 6-2 Sample calibration tags – T-Series sensors
Density calibration factors
If your sensor tag does not show a D1 or D2 value:
For D1, enter the Dens A or D1value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air.
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.
If your sensor tag does not show a K1 or K2 value:
For K1, enter the first 5 digits of the density calibration factor. In the sample tag in Figure 6-1,
this value is shown as
12500.
For K2, enter the second 5 digits of the density calibration factor. In the sample tag in
Figure 6-1, this value is shown as
14286.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration
factor. In the sample tag in Figure 6-1, this value is shown as
4.44.
Newer tag Older tag
19.0005.13
19.0005.13
0.0010
0.9980
12502.000
14282.000
4.44000
310
12500142864.44
12500142864.44
Newer tag Older tag
Zobrazit stránku 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 263 264

Komentáře k této Příručce

Žádné komentáře