70 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Using the Transmitter continued
Record the following process variables:
• Flow rate
• Density
•Temperature
• Tube frequency
• Pickoff voltage
•Drive gain
For information on using this information in troubleshooting, see Section 10.12.
7.4 Viewing process variables
Process variables include measurements such as mass flow rate, volume flow rate, mass total, volume
total, temperature, and density.
You can view process variables with the display (Model 1700 and 2700 transmitters only), ProLink II,
or the Communicator.
7.4.1 With the display
The display reports the abbreviated name of the process variable (e.g.,
DENS for density), the current
value of that process variable, and the associated unit of measure (e.g., G/CM3). See Appendix J for
information on the codes and abbreviations used for display variables.
The display must be configured. For instructions on how to configure the display, see Section 8.14.1.
To view a process variable with the display, press
Scroll until the name of the desired process variable
either:
• Appears on the process variable line, or
• Begins to alternate with the units of measure
See Figure 2-1.
7.4.2 With ProLink II
To view process variables with ProLink II software:
1. The
Process Variables window opens automatically when you first connect to the transmitter.
2. If you have closed the
Process Variables window:
a. Open the
Prolink menu.
b. Select
Process Variables.
7.4.3 With a Communicator
To view process variables with a Communicator:
1. Press
1, 1.
2. Scroll through the list of process variables by pressing
Down Arrow.
3. Press the number corresponding to the process variable you wish to view, or highlight the
process variable in the list and press
Right Arrow.
Komentáře k této Příručce