72 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Using the Transmitter continued
7.6.1 Using the status LED
For these transmitters, the status LED is located on the front panel. This LED shows
transmitter status as described in Table 7-1.
7.6.2 Using the display
The display reports alarms in two ways:
With the status LED, which reports only that one or more alarms has
occurred
Through the alarm queue, which reports each specific alarm
Note: If access to the alarm menu from the display has been disabled (see
Section 8.14.1), then the display will not list alarm codes in an alarm queue, and the
status LED will not flash. It will indicate status using solid green, yellow, or red.
For these transmitters, the status LED is located at the top of the display (see Figure 7-1). It can be in
one of six possible states, as listed in Table 7-1.
Figure 7-1 Display status LED
Table 7-1 Transmitter status reported by the Model 1500/2500 status LED
Status LED state Alarm priority Definition
Green No alarm Normal operating mode
Flashing yellow No alarm Zero in progress
Yellow Low severity alarm Alarm condition: will not cause measurement error
Outputs continue to report process data
Red High severity (critical fault) alarm Alarm condition: will cause measurement error
Outputs go to configured fault indicators
Model 1500 AN
Model 2500 CIO
Model 1700 AN
Model 1700 IS
Model 2700 AN
Model 2700 IS
Model 2700 CIO
Status LED