Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 29
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Chapter 5
Flowmeter Startup
5.1 Overview
This chapter describes the procedures you should perform the first time you start the flowmeter. You
do not need to use these procedures every time you cycle power to the flowmeter.
The following procedures are discussed:
Applying power to the flowmeter (see Section 5.2)
Zeroing the flowmeter (see Section 5.5)
Performing a loop test on the transmitter outputs (see Section 5.3)
Trimming the mA outputs (see Section 5.4)
Figure 5-1 provides an overview of the flowmeter startup procedures.
Figure 5-1 Startup procedures
Note: All ProLink II procedures provided in this chapter assume that your computer is already
connected to the transmitter and you have established communication. All ProLink II procedures also
assume that you are complying with all applicable safety requirements. See Chapter 3 for more
information.
Note: If you are using AMS, the AMS interface will be similar to the ProLink II interface described in
this chapter.
Note: All Communicator procedures provided in this chapter assume that you are starting from the
“Online” menu. See Chapter 4 for more information.
Finish
Start
Apply power
Perform a
loop test
Tr i m m A
outputs (if
necessary)
Zero the
flowmeter