
38 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Flowmeter Startup continued
5.5 Zeroing the flowmeter
Zeroing the flowmeter establishes the flowmeter’s point of reference when there is no flow.
When you zero the flowmeter, you may need to adjust the zero time parameter. Zero time is the
amount of time the transmitter takes to determine its zero-flow reference point. The default zero time
is 20 seconds.
•A long zero time may produce a more accurate zero reference but is more likely to result in a
zero failure. This is due to the increased possibility of noisy flow, which causes incorrect
calibration.
•A short zero time is less likely to result in a zero failure but may produce a less accurate zero
reference.
For most applications, the default zero time is appropriate.
Note: In some menus, a convergence limit parameter is displayed. Micro Motion recommends that you
use the default value for convergence limit.
Note: Do not zero the flowmeter if a high severity alarm is active. Correct the problem, then zero the
flowmeter. You may zero the flowmeter if a low severity alarm is active. See Section 7.6 for
information on viewing transmitter status and alarms.
You can zero the flowmeter with the display (if the transmitter has a display), ProLink II, the
Communicator, or with the Zero button on the transmitter (Model 1500/2500 transmitters only).
If the zero procedure fails, see Section 10.6 for troubleshooting information.
5.5.1 Preparing for zero
To prepare for the zero procedure:
1. Apply power to the flowmeter. Allow the flowmeter to warm up for approximately 20 minutes.
2. Run the process fluid through the sensor until the sensor temperature reaches the normal
process operating temperature.
3. Close the shutoff valve downstream from the sensor.
4. Ensure that the sensor is completely filled with fluid.
5. Ensure that the process flow has completely stopped.
CAUTION
If fluid is flowing through the sensor, the sensor zero
calibration may be inaccurate, resulting in inaccurate
process measurement.
To improve the sensor zero calibration and measurement
accuracy, ensure that process flow through the sensor
has completely stopped.
Komentáře k této Příručce