Emerson 3051S Specifikace Strana 128

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 240
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 127
Reference Manual
00809-0200-4801, Rev CA
July 2010
Rosemount 3051S
A-18
SPM3_BLOCK_
TYPE
48 All Function
Blocks in the
device
0 N/A The function block type for the SPM3 variable
SPM3_BLOCK_
TAG
49 All Function
Blocks in the
device (see
Table 33)
NULL N/A Block tag for the SPM3 process variable
SPM3_PARAM_
INDEX
50 All DS-65
parameters for
the block type of
the entered
block tag. Any
number can be
entered, but an
invalid index will
result in a
rejection of a
“True” write to
the
SPM_ACTIVE
parameter
0 N/A OD Parameter index for the SPM3 Block Tag that is entered.
SPM3_STATUS 51 1.Inactive
2.Learning
3.Verifying
4.No Detections
5.Mean Change
Detected
6.High Variation
Detected
7.Low Dynamics
Detected
8.Not Licensed
Inactive N/A Read
Only
Status of the SPM3 Statistical Process Monitoring
SPM3_
TIMESTAMP
52 All Times and
Dates
0 time Read
Only
Timestamp of last SPM3 Statistical Process Monitoring status
change
SPM3_USER_
COMMAND
53 1.Detect
2.Learn
3.Quit
4.Idle
4 N/A User control for the SPM3 Statistical Process Monitoring
session.
SPM3_MEAN 54 0.0f N/A Read
Only
Last Mean of SPM3
SPM3_MEAN_
CHANGE
55 0.0f % Read
Only
% Change in SPM3 Mean with respect to Baseline Mean
SPM3_STDEV 56 0.0f N/A Read
Only
Last Stdev of SPM3
SPM3_STDEV_
CHANGE
57 0.0f % Read
Only
Change in SPM3 Stdev with respect to Baseline Stdev
SPM3_BASELIN
E_MEAN
58 0.0f N/A Read
Only
Baseline Mean for SPM3
SPM3_BASELIN
E_STDEV
59 0.0f N/A Read
Only
Baseline Stdev for SPM3
SPM3_MEAN_
LIM
60 0.0f 0.0f % % Change in mean for SPM3 allowed by user
SPM3_HIGH_
VARIATION_LIM
61 0.0f 0.0f % % Increase in dynamics for SPM3 allowed by user
SPM3_LOW_
DYNAMICS_LIM
62 0.0f 0.0f % % Decrease in dynamics for SPM3 allowed by user
SPM4_BLOCK_
TYPE
63 All Function
Blocks in the
device
0 N/A The function block type for the SPM variable
Parameter Index Valid Range Initial Value Units Other Description
Zobrazit stránku 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 239 240

Komentáře k této Příručce

Žádné komentáře