Reference Manual
00809-0200-4801, Rev CA
July 2010
B-13
Rosemount 3051S
Advanced HART Diagnostics Suite
(Option Code DA2)
The 3051S provides Abnormal Situation Prevention indication for a breakthrough in
diagnostic capability. The 3051S ASP Diagnostics Suite for HART includes Statistical
Process Monitoring (SPM), variable logging with time stamp and advanced process alerts.
The enhanced EDDL graphic display provides an intuitive and user-friendly interface to better
visualize these diagnostics.
The integral SPM technology calculates the mean and standard deviation of the process
variable 22 times per second and makes them available to the user. The 3051S uses these
values and highly flexible configuration options for customization to detect many user-defined
or application specific abnormal situations (e.g. detecting plugged impulse lines and fluid
composition change). Variable logging with time stamp and advanced process alerts capture
valuable process and sensor data to enable quick troubleshooting of application and
installation issues.
FOUNDATION fieldbus
Power Supply
External power supply required; transmitters operate on 9.0 to 32.0 Vdc transmitter terminal
voltage.
Current Draw
17.5 mA for all configurations (including LCD display option)
F
OUNDATION fieldbus Parameters
Standard Function Blocks
Resource Block
• Contains hardware, electronics, and diagnostic information.
Transducer Block
• Contains actual sensor measurement data including the sensor diagnostics and the ability
to trim the pressure sensor or recall factory defaults.
LCD Block
• Configures the local display.
2 Analog Input Blocks
• Processes the measurements for input into other function blocks. The output value is in
engineering or custom units and contains a status indicating measurement quality.
PID Block with Auto-tune
• Contains all logic to perform PID control in the field including cascade and feedforward.
Auto-tune capability allows for superior tuning for optimized control performance.
3051SMV and 3051SF_1-7
3051S and 3051SF_D with HART Diagnostics
(option code DA1)
Maximum Loop Resistance = 43.5 * (Power Supply Voltage – 12.0)
The HART communicator requires a minimum
loop resistance of 250
for communication.
Schedule Entries 14 (max.)
Links 30 (max.)
Virtual Communications Relationships (VCR) 20 (max.)
Voltage (Vdc)
Load (Ohms)
Operating
Region
1322
1000
500
0
12.0 20 30
42.4