Emerson 3051S Specifikace Strana 142

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 240
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 141
Reference Manual
00809-0200-4801, Rev CA
July 2010
Rosemount 3051S
B-12
Process Temperature RTD Interface
(3051SMV_ _1 or 3, 3051SF_1, 3, 5 or 7)
Minimum Span = 50 °F (28 °C)
Service
3051S, 3051SMV_P, and 3051SF_5, 6, 7, or D (Direct Process Variable Output):
Liquid, gas, and vapor applications
3051SMV_M and 3051SF_1, 2, 3, or 4 (Mass and Energy Flow Output):
Some fluid types are only supported by certain measurement types
4–20 mA/HART
Zero and Span Adjustment
Zero and span values can be set anywhere within the range.
Span must be greater than or equal to the minimum span.
Output
Two-wire 4–20 mA is user-selectable for linear or square root output. Digital process variable
superimposed on 4–20 mA signal, available to any host that conforms to the HART protocol.
Power Supply
External power supply required.
3051S and 3051SF_D: 10.5 to 42.4 Vdc with no load
3051S and 3051SF_D with Advanced HART Diagnostics Suite: 12 to 42.4 Vdc with no load
3051SMV and 3051SF_1-7: 12 to 42.4 Vdc with no load
Load Limitations
Maximum loop resistance is determined by the voltage level of the external power supply, as
described by:
Fluid Compatibility with Pressure and Temperature Compensation
Available — Not available
Ordering
Code
Fluid Types
Measurement Type Liquids Saturated Steam Superheated Steam Gas and Natural Gas
1 DP / P/ T (Full Compensation)
2 DP / P
3 DP / T
4 DP only
3051S and 3051SF_D
Maximum Loop Resistance = 43.5 * (Power Supply Voltage – 10.5)
The HART communicator requires a minimum
loop resistance of 250
for communication.
Voltage (Vdc)
Load (Ohms)
Operating
Region
1387
1000
500
0
10.5 20 30
42.4
Zobrazit stránku 141