Reference Manual
00809-0200-4801, Rev CA
July 2010
Rosemount 3051S Series
2-12
1. Loosen the housing rotation set screw.
2. First rotate the housing clockwise to the desired location. If the desired
location cannot be achieved due to thread limit, rotate the housing
counter clockwise to the desired location (up to 360° from thread limit).
3. Retighten the housing rotation set screw.
In addition to housing rotation, the optional LCD display can be rotated in
90-degree increments by squeezing the two tabs, pulling out, rotating and
snapping back into place.
NOTE
If LCD pins are inadvertently removed from the interface board, carefully
re-insert the pins before snapping the LCD display back into place.
Configure Security The 3051S FOUNDATION fieldbus transmitter has a hierarchy of security. The
SECURITY switch located on the electronics provides the highest level of
security. In the ON position, all writes to the transmitter are disabled.
Simulate The SIMULATE switch is located on the Electronics. It is used in conjunction
with the transmitter simulate software to simulate process variables and/or
alerts and alarms. To simulate variables and/or alerts and alarms, the
SIMULATE switch must be moved to the ENABLE position and the software
enabled through the host. To disable simulation, the switch must be in the
DISABLE position.
Komentáře k této Příručce